Home › Guides › USP 88 December 1, 2026: Class VI transition tracker
USP <88> December 1, 2026: USP Class VI transition tracker
By The MedResin team · Last updated · Data: 417 public grades from 29 manufacturers, checked 2026-10-20
Revised USP <88> becomes official on December 1, 2026.
Date from the USP-NF notice.
What changes on December 1, 2026
- Class I to VI is removed. The revised chapter <88> Biological Reactivity Tests, In Vivo drops the classification of plastics into Classes I to VI, so USP no longer defines a "Class VI" designation (USP-NF notice, revised <88>).
- One in vivo test remains. Of the systemic injection, intracutaneous and implantation tests in the current chapter, only the systemic injection test remains.
- <87> changes the same day. USP <87> (in vitro biological reactivity) is revised and expanded on the same date (revised <87>).
- Scope is drug packaging. Both revised chapters are written for packaging systems for drugs and for packaging of combination products regulated by FDA's drug center (CDER).
- Devices point to FDA guidance on ISO 10993-1. For medical devices, and combination products regulated as devices, both chapters refer readers to FDA's device guidance, which is built on ISO 10993-1 (FDA ISO 10993-1 guidance). USP does not present ISO 10993 as a replacement for Class VI.
- Existing Class VI data stays as historical evidence. The change does not invalidate testing already performed (Saint-Gobain; see also ISPE). What ends is the designation for new claims.
Where MedResin's 417 grades stand
Each grade is counted once, by what its cited manufacturer source states. ISO 10993 follows MedResin's counting rule: cytotoxicity-only statements are not counted as ISO 10993.
| Status in the cited source | Grades | Share of 417 |
|---|---|---|
| USP Class VI and ISO 10993 | 182 | 44% |
| USP Class VI without a counted ISO 10993 statement | 80 | 19% |
| of which cite cytotoxicity testing only | 53 | |
| of which report a failed ISO 10993 endpoint | 1 | |
| of which have no ISO 10993 or cytotoxicity mention | 26 | |
| ISO 10993 without USP Class VI | 41 | 10% |
| Neither stated (includes 6 with a limited ISO 10993 statement only) | 114 | 27% |
The 80 "USP VI only" grades, from 13 of the 24 manufacturers that state USP Class VI, are the ones where Class VI is the main biocompatibility claim in the cited source. Blank means "not stated in the cited source", not "not tested". Suppliers may hold further data. Polymer-group detail is in the USP Class VI ends data report.
By manufacturer
Sorted by number of USP Class VI grades. Manufacturer names link to their full grade lists.
| Manufacturer | Grades | USP VI + ISO 10993 | USP VI only | ISO 10993 only | Neither stated |
|---|---|---|---|---|---|
| Covestro | 46 | 40 | 6 | ||
| Evonik | 39 | 38 | 1 | ||
| Celanese | 28 | 28 | |||
| INEOS Styrolution | 25 | 23 | 2 | ||
| Arkema | 18 | 18 | |||
| INVISTA | 13 | 13 | |||
| Wacker | 12 | 12 | |||
| Teknor Apex | 27 | 6 | 5 | 16 | |
| LyondellBasell | 15 | 9 | 6 | ||
| Mitsubishi Chemical Advanced Materials | 9 | 8 | 1 | ||
| Momentive | 9 | 9 | |||
| Lubrizol | 64 | 8 | 56 | ||
| Eastman | 13 | 8 | 1 | 4 | |
| ExxonMobil | 7 | 7 | |||
| BASF | 10 | 6 | 4 | ||
| KRAIBURG TPE | 6 | 4 | 1 | 1 | |
| Zeon | 4 | 4 | |||
| Chevron Phillips Chemical | 3 | 3 | |||
| Plaskolite | 3 | 3 | |||
| Syensqo (Solvay) | 5 | 1 | 1 | 3 | |
| TOPAS Advanced Polymers | 5 | 2 | 3 | ||
| Ensinger | 3 | 2 | 1 | ||
| TotalEnergies | 3 | 1 | 2 | ||
| Victrex | 1 | 1 | |||
| Borealis | 26 | 14 | 12 | ||
| SABIC | 13 | 13 | |||
| Delrin (formerly DuPont) | 6 | 6 | |||
| NuSil (Avantor) | 3 | 3 | |||
| Invibio (Victrex) | 1 | 1 | |||
| All | 417 | 182 | 80 | 41 | 114 |
By resin family
Families as listed in MedResin, sorted by number of USP Class VI grades. Small families describe only the grades in MedResin, not the market.
The 80 USP VI only grades
Grades whose cited source states USP Class VI but no counted ISO 10993 statement, by manufacturer. These are the entries to check with the supplier first.
Cytotoxicity only or reported ISO failure (54)
BASF (6): Ultradur PRO B4521 · Ultraform N2320 003 PRO AT · Ultraform S1320 003 PRO AT · Ultraform S2320 003 PRO AT · Ultraform S2320 003 PRO TR AT · Ultraform W2320 003 PRO AT
Celanese (28): Celanex 2401MT · Celanex 2402MT · Celanex 2404MT · Celanex 2405MT · Fortron MT 9140L4 · Fortron MT 9140L6 · GUR 1001 · GUR 1020 · GUR 1020-E · GUR 1050 · GUR 1050-E · Hostaform MT 12R01 · Hostaform MT 12U01 · Hostaform MT 12U03 · Hostaform MT 24F01 · Hostaform MT 24U01 · Hostaform MT 8F01 · Hostaform MT 8F02 · Hostaform MT 8R02 · Hostaform MT 8U01 · Hostaform MT 8U05 · Hostaform MT SlideX 1203 · Hostaform MT SlideX 2404 · Vectra MT 1300 · Vectra MT 1310 · Vectra MT 1335 · Vectra MT 4310 · Vectra MT 4350
Evonik (1): VESTAMID Care ME26
INVISTA (13): 11M35V · 11R12A · 12G50M · 13R9A · 13T25A · 23M2A · 23N10A · 23R2A · 23R2B · H0500HN · P4C5N-046 · P5M6K-080 · P9M7R-056
Mitsubishi Chemical Advanced Materials (1): Proteus LSG HP PP-H
Teknor Apex (5): Medalist MD-91255 · Medalist MD-91760 · Medalist MD-91970 · Medalist MD-92260 · Medalist MD-96267
No ISO 10993 or cytotoxicity mention (26)
Chevron Phillips Chemical (3): Marlex 9012 · Marlex 9018 · Marlex HHM 5502BN
Lubrizol (8): Pellethane 2363-55D · Pellethane 2363-55DE · Pellethane 2363-65D · Pellethane 2363-75D · Pellethane 2363-80A · Pellethane 2363-80AE · Pellethane 2363-90A · Pellethane 2363-90AE
LyondellBasell (9): Purell 1840H · Purell 3020D · Purell 3220D · Purell PD702 · Purell PE53003 · Purell PE53003Z · Purell PF511 · Purell PF531 · Purell SR549M
Plaskolite (3): OPTIX CA-1000IG · OPTIX CA-927 G · OPTIX CA-927 GHF
Syensqo (Solvay) (1): Udel P-1700
TotalEnergies (1): 3847MR
Victrex (1): VICTREX PC (family)
Tracker log
- 2026-11-02: Tracker published, with counts by manufacturer and family from the current data.
- 2026-10-23: Data report on which USP Class VI resins also state ISO 10993.
- 2026-10-17: USP <88> section added to the USP Class VI vs ISO 10993 guide.
Counts on this page are recomputed at every site build. Changes to individual grades are logged on the changelog.
Common questions
When does the revised USP <88> become official?
December 1, 2026. USP's notice gives that date as the official date of revised general chapter <88> Biological Reactivity Tests, In Vivo (USP-NF notice: https://www.uspnf.com/notices-gc-88-dependencies-20240426).
Is USP Class VI being removed?
Yes, as a USP designation. The revised <88> removes the chapter's classification of plastics into Classes I to VI, so from December 1, 2026 USP no longer defines a "Class VI". Of the three in vivo tests in the current chapter, only the systemic injection test remains.
Is existing USP Class VI test data still valid?
The change does not invalidate Class VI testing already performed. Industry commentary (Saint-Gobain, ISPE) describes it as historical evidence that can still support a material's safety documentation where appropriate. What ends is USP's Class VI designation for new claims.
What replaces USP Class VI for medical devices?
USP does not present ISO 10993 as a replacement. The revised <88> and <87> are written for drug packaging and for combination products regulated by FDA's drug center (CDER). For medical devices they refer readers to FDA's device guidance, which is built on the ISO 10993-1 biological evaluation framework. ISO 10993-1 was already the device framework.
Does the revised USP <88> apply to medical devices?
Its scope is packaging systems for drugs and packaging of combination products regulated by CDER. For devices, and combination products regulated as devices, it refers readers to FDA's device guidance.
How many medical resins state USP Class VI but no ISO 10993?
In MedResin, 80 of the 262 grades that state USP Class VI have no counted ISO 10993 statement in the cited supplier source (53 cite cytotoxicity only, 1 reports a failed ISO 10993 endpoint and 26 have no ISO 10993 mention). 182 state both. Counts update with each site build.
Will MedResin remove USP Class VI from grade pages?
No. Each entry records what the supplier stated about testing done under the chapter as it stood. We keep those statements as stated and update them as suppliers revise their documents.
What should I ask my supplier before December 1, 2026?
Ask what biocompatibility data it will provide after December 1, 2026 (for a device, typically ISO 10993 test reports; USP <87> data where relevant), whether the data covers your exact grade and color, and whether existing Class VI letters will be reissued or withdrawn. Base the device evaluation on ISO 10993-1.
Browse the grades
All 262 USP Class VI grades · All ISO 10993 grades · USP Class VI vs ISO 10993 · All manufacturers · All resin families · Compare grades · Download the data
Sources: USP-NF notice (official date December 1, 2026) · revised <88> (USP DOI) · revised <87> (USP DOI) · FDA ISO 10993-1 guidance · Saint-Gobain · ISPE. Last updated 2026-11-02.
Free newsletter: new grades and change alerts. Read the latest issue
Grade data last checked 2026-10-20. Blank cells mean not stated by the source. Each grade page cites its source. About the data · Suggest a correction
More guides: Which medical plastics survive gamma sterilization? · Autoclave-safe resins compared · USP Class VI vs ISO 10993 · EtO-compatible materials for catheters and tubing · Choosing a TPU for medical tubing · USP Class VI ends Dec 1: which resins also have ISO 10993 data? · All guides
Overview of every polymer group: Medical-grade plastics: grades by polymer and sterilization